Download Manual
Transcript
THERAPY 5 Select YES to enter Manual mode. Manual Mode Enter Manual Mode? Yes No Push Speed Dial to confirm • AED/Passcode—A passcode screen appears: Manual Mode Rotate and press the SPEED DIAL to enter the passcode. Enter Manual Mode passcode The code changes to dots to protect the passcode, and the defibrillator enters Manual mode. You have three opportunities to enter the correct password. After an incorrect attempt, the message INCORRECT--TRY AGAIN appears. After three incorrect attempts, the message ACCESS DENIED appears, and the defibrillator returns to AED mode. • Restricted—A MANUAL MODE DISABLED message appears, an alert tone sounds, and the LIFEPAK 15 monitor/defibrillator returns to AED mode. It is important that all users of the LIFEPAK 15 monitor/defibrillator be thoroughly familiar with the monitor/defibrillator settings and operation before use. Special AED Setup Options The following descriptions of AED prompts (voice and text) explain special setup options. Initial CPR - CPR First When the INITIAL CPR option is set to CPR FIRST, you are prompted to START CPR immediately after the AED is turned on, and before an analysis. ©2007-2010 Physio-Control, Inc. LIFEPAK 15 Monitor/Defibrillator Operating Instructions 5-15