Download Manual
Transcript
Monitoring the ECG Selecting ECG Lead The LIFEPAK 15 monitor/defibrillator includes two methods for selecting or changing the ECG lead. To select or change the displayed ECG lead using the LEAD button: : 1. Press LEAD. If any ECG lead currently appears on the Home Screen, the lead changes to PADDLES. If PADDLES lead is currently displayed, the lead changes to Lead II. Lead Paddles I II III AVR AVL AVF II, CO2 I, II, III AVR, AVL, AVF 2. While the LEAD menu is displayed, press LEAD again or rotate the SPEED DIAL to the desired lead. Note: If lead sets are predefined for Channels 2 and 3, the lead sets show on the menu. The ECG cable that is connected to the device, such as 3lead or 5-wire, determines the leads you can select. For information about defining lead sets, see the LIFEPAK 15 Monitor/Defibrillator Setup Options provided with your device. To select or change the displayed ECG lead using the SPEED DIAL: 1. For the primary ECG, outline and select CHANNEL 1 and then select LEAD. Channel 1 Lead Size 1.0 2. Rotate the SPEED DIAL to the desired ECG lead. 3. Press the SPEED DIAL to select the ECG lead. 4. Repeat this procedure to select or change displayed ECG waveforms for Channels 2 and 3. Note: The ECG shows dashed lines until the electrodes are connected to the patient. Note: When the VF/VT ALARM is ON, you are limited to PADDLES lead or Lead II in Channel 1. See "Setting Alarms" on page 3-21. 4-4 LIFEPAK 15 Monitor/Defibrillator Operating Instructions