DATA MANAGEMENT 7 Table 7-3 Waveform Events (Continued) EVENT NAME WAVEFORM DATA (WHEN CAPTURED) SHOCK X 3 seconds prior to shock and 5 seconds after shock PACING X STARTED 8 seconds prior to increase of current from 0 PACING X SET 8 seconds after ppm and mA are stable for 10 seconds PACING X CHANGED 8 seconds after pacing rate, current, or mode is changed PACING X STOPPED 3 seconds prior to pacing current is zero and 5 seconds after PACING X PAUSED Initial 8 seconds while PAUSE is pressed ALARM* 3 seconds prior to violated parameter and 5 seconds after EVENT* 3 seconds prior to event selection and 5 seconds after PRINT 3 seconds prior to pressing PRINT and 5 seconds after 12-LEAD 10 seconds after 12-LEAD is pressed SNAPSHOT 3 seconds prior to and 5 seconds after SNAPSHOT requested VITAL SIGNS 3 seconds prior to and 5 seconds after vital signs are acquired *To reduce the length of the CODE SUMMARY report, storing waveform data with these events can be set to OFF (see the LIFEPAK 15 Monitor/Defibrillator Setup Options provided with your device). Waveform events are preceded by a header that includes the following information: • Patient data • Vital signs • Event name • Device configuration information • Therapy data* *Patient impedance (in ohms) appears on shock reports when using disposable defibrillation electrodes. This impedance is measured just prior to the shock and is used to determine voltage compensation. Figure 7-3 shows four examples of waveform events as they would appear in the CODE SUMMARY report. ©2007-2010 Physio-Control, Inc. LIFEPAK 15 Monitor/Defibrillator Operating Instructions 7-7