Download Specifying Calibration Standards and Kits for Agilent Vector Network
Transcript
Connector definitions In addition to calibration standard definitions and standard class assignments, calibration kits also provide definitions of connectors. Agilent’s PNA network analyzer family uses the connector definition to define connector (Figure 2): • • • • • • Frequency Range Gender (male, female, no gender) Impedance Media (coax, waveguide, etc) Cutoff Frequency (waveguide) Height/Width Ratio (waveguide) Figure 2. PNA connector entry screen A calibration kit may be defined with multiple connectors. Each 1-port calibration standard must be associated with a defined connector. Two port standards, such as thrus and adapters, may be associated with two different defined connectors. 5