Download - Transcat
Transcript
Fluke 190-062, -102, -104, -202, -204, -502, -504 Service Manual Input B Trigger Sensitivity Test Proceed as follows to test the Input B trigger sensitivity: 1. Connect the Test Tool to the 5502A as shown in Figure 4-8. PM9091 PM9091 50 OHM e.g. PM9585 or TRM50 50 OHM e.g. PM9585 or TRM50 For 190-502 and -504 use only TRM50! A METER USE 50 OHM TERMINATION B B A C D 4 CHANNELS 2 CHANN. + METER NORMAL SCOPE FLUKE 5502A CALIBRATOR perf-ver-e2.eps Figure 4-8. 5502A Scope Output to Test Tool Input B 2. Reset the Test Tool to select the Test Tool setup: B 3. Press 4. Press to turn Input B on. TRIGGER mV and use F1 to select Input B as trigger source. RANGE 5. Use RANGE and V to change the sensitivity range to select manual sensitivity ranging, and lock the Input B sensitivity range on 2 V/div. 6. Use s TIME ns to select the time base listed in Table 4-11. 7. Set the 5502A to source the leveled sine wave given in the first row of Table 4-10. 8. Adjust the 5502A output voltage until the displayed trace has the amplitude indicated under the appropriate column of Table 4-11. 9. Verify that the signal is well triggered. to enable for manual Trigger If not, press TRIGGER and use F3 Level adjustment. Adjust the trigger level and verify that the signal will be triggered now. The trigger icon ( ) indicates the trigger level. 10. Continue through the test points. 11. When you are finished, set the 5502A to Standby. 4-24