Download SE Seires
Transcript
COMMAND
SM?
SBI n
SBI?
SAS n
SAS?
NAME
VALUE
0=True RMS, 1=Average
n=0-4
0=Off, 1=Ser/Prod, 2=Serial#,
3=Product#, 4=Run File
n=0-1
0=Off, 1=On
Barcode IP Format
Auto Start Format
8.4.6 Query Commands
These query commands will retrieve data from the instrument. The GPIB bus
application requires an IEEE-488 read command to be sent after the query command.
These commands include functions for retrieving test data, test results and remote
hardware status as well as setup file information.
COMMAND
TD?
RDM?
NAME
List Testing Data
Read Max Current
VALUE
Test In Process
ACW: Total current,
Real Current DCW:
current
step number = 1-200
RD <step
number>?
RR?
RI?
RS?
List Results Data
RU?
FR nnnn
LF?
Read USB Stick
File Rename
List File Name
LF nn?
LP?
LP <step
number>?
LS2?
LS2 <step
number>?
List File Name
List Prompt
List Prompt by step number
Read Remote Reset
Read Remote Interlock
Read Scanner Status
List Step Parameters
List Step Parameters by step
number
1=Open, 0=Closed
1=Open, 0=Closed
0=None, 1=Int, 2=Ext
1, 3=Int+Ext 1,
4=Ext 2, 5=Int+Ext 2,
6=Ext 1+Ext 2,
7=Int+Ext 1+Ext 2
0=NO, 1=YES
nnn=name
Active selected
memory
Active selected file
Active selected Step
step number = 1-200
step number = 1-200
TD?
Read the active data being displayed on the LCD display while the test is in process.
Will also read the last data taken when the test sequence has completed. Each
parameter is separated by commas and includes step number, test type, test status,
and metering. The syntax for this command response is {step, test type, status, meter
1, meter 2, meter 3}. ACW test displays 4 meters. Each meter will contain only the
value and not the units. In the case of DCW current where both uA and mA are used
79