Download Owner / Calibration Manual
Transcript
Model 8903B Operation < START 89038 QNTERRUP; I SET TRIGGER MODE TO HOLD (HP-I8 CODE T1) ENABLE SRQ ON DATA READY (HP-16 CODE 22.3SP) 1 TO OTHER SERVICE SERVICE INSTRUMENT ROUTINES ROUTINES No ON 1 SRO) -1 READ STATUS BYTE FROM 8 9 0 3 8 1 A SA0 FROM 89038 ~ CONF 1 GURE MEASUREMENT TEST OTHER 8 9 0 3 8 SRO CONDITIONS I I TRIGGER MEASUREMENT (HP-16 CODE T 2 OR T 3 ) . No I I PROCESS OTHER ROUTINES UNTIL SRQ CAUSES INTERRUPT DUE TO DATA READ DATA FROM 8 9 0 3 8 A I RETURN Figure 3-4. Example Flow Chart for Driving the Audio Analyzer Using the Require Service Message (SRQ 3-7. HP-IB SYNTAX AND CHARACTERISTICS SUMMARY Address: Set in binary by internal switches - may be displayed on front panel using Special Function 21, HP-IB Address. (Factory set to 28 decimal; 11100 binary.) General Operating Syntax: (Excluding Rapid Frequency Count and Rapid Source modes.)* [Automatic Operation] [SourceFrequency] [Source Amplitude] [Measurement] [Filters][SpecialFunctions] [Log/Lin] [Ratio]... ...[Start Frequency] [Stop Frequency] [Plot Limit] [Sweep] [Trigger] Numeric Data Input Format: (Except in Rapid Source mode.)* +DDDDDE+NN 5-Digit Signed Mantissa (leading decimal not allowed) Exponent Magnitude Exponent Sign Indicates Exponent Follows *For information on Rapid Frequency Count or Rapid Source modes refer to them by name in the Detailed Operating Instructions. 3-35