Download 793 Tuning and Calibration
Transcript
Force Sensor Calibration You must know which DC conditioner is associated with the force sensor. You also need to know which of the two DC conditioners on the module is used by your load cell. 497.22 DC Cond C h C h U Lim Excit Fail R-Cal jack for the sensor connected to Jx01. The shunt resistors for the Model 497.22 DC Conditioner are packaged in a thumb-nail sized module. It makes it easier to install and remove the shunt calibration resistor. RCal R-Cal jack for the sensor connected to Jx02. RCal Out 1 Out 10. Verify that force is still zero. While it is unlikely, it is possible for the force signal to change when the control mode changes. If it does: Standalone FlexTest SE Controllers: Click Auto Offset on the Offset/Zero submenu to zero the force output. Select Setup > Force > Offset/Zero > Auto Offset Automated Controllers: Click Auto Offset on the Offset/Zero tab (Inputs panel) to zero the force output. 11. Perform shunt calibration. Standalone FlexTest SE Controllers: Select Setup > Force > Calibration > Select Cal Type A. Select required Shunt Display units. B. Ensure that Shunt State (+) is set to On. Polarity is always positive. Note the Shunt Ref (+) value, it should be 60%–90% (80% is ideal) of the calibrated range of the sensor. C. Click Update to copy the current shunt reference value into the Shunt Reference box. D. Set Shunt State (+) to Off. MTS 793 Tuning and Calibration 181