Download MAC2000 Operators Manual
Transcript
System Configuration Resting ECG Setup Fields–Page 3 Field Description Lead Sequence Determines the lead sequence to use. Values are: • Standard • Cabrera • NEHB • SEQ4 SEQ4 allows you to configure a custom 12-lead sequence using the following fields. If either 12SL option (ME12 or MI12) is activated, you must select leads I (-I), II (-II), V1, V2, V3, V4, V5, and V6 for a correct 12SL analysis. 118 Sequence Name Set the display name for a custom lead sequence. Available only if SEQ4 is selected for the Lead Sequence. 1–12 Lead Twelve fields that allow you to define the sequence in which the leads are displayed. Available only if SEQ4 is selected for the Lead Sequence. 1–12 Label Twelve fields that allow you to define the labels that are displayed/printed for the corresponding leads. Available only if SEQ4 is selected for the Lead Sequence. 1–6 Rhythm Leads Six fields that allow you to define the rhythm leads and their sequence. You can select the rhythm leads for all four lead sequences. MAC™ 2000 ECG Analysis System 2053535-002C