5500A
Operator Manual
7-10. Resistance Accuracy
The Resistance Accuracy test verifies the accuracy of synthesized resistance at the
5500A Calibrator front panel NORMAL terminals. For resistances of less than 110 kΩ,
use the 4-wire COMP option. For resistances of 110 kΩ or higher, the COMP option is
automatically turned off.
Nominal Value (Ω)
7-10
Measured Value (Ω)
Deviation %
90-Day Spec. (mΩ or %)
0Ω
6 mΩ
2Ω
0.309%
10.9 Ω
0.064%
11.9 Ω
0.135%
19 Ω
0.088%
30 Ω
0.059%
33 Ω
0.052%
109 Ω
0.021%
119 Ω
0.020%
190 Ω
0.015%
300 Ω
0.012%
330 Ω
0.025%
1.09 kΩ
0.012%
1.19 kΩ
0.012%
1.9 kΩ
0.010%
3 kΩ
0.009%
3.3 kΩ
0.025%
10.9 kΩ
0.012%
11.9 kΩ
0.012%
19 kΩ
0.010%
30 kΩ
0.009%
33 kΩ
0.026%
109 kΩ
0.013%
119 kΩ
0.014%
190 kΩ
0.012%
300 kΩ
0.011%
330 kΩ
0.028%
1.09 MΩ
0.016%
1.19 MΩ
0.016%
1.9 MΩ
0.014%
3 MΩ
0.013%
3.3 MΩ
0.062%