Reference Overview To take a common mode or differential TDR measurement (cont.) Control elements & resources 20. When the TDR steps on the two channels are opposite (one positive and one negative), you can define a math waveform that represents the difference signal by pressing the VERTICAL MENU button, selecting the Vert tab, selecting Waveform M1, On, and then selecting Define, C1, +, C2, Math Waveform On, and OK. Set the scale to ρ. (If using volts, subtract the waveforms.) Take a 21. Take your measurement. For more information see Take measurement automatic measurements on page 25, or Take cursor measurements on page 26. TDT 22. You can make forward and reverse Time Domain measurements Transmission (TDT) measurements using the 80E04. To perform a TDT measurement: connect one sampling module channel to the input of the device under test and the other sampling module channel to the output of the device under test. Device under test 23. Then alternately enable the step generators on one channel while sampling the transmitted signal on the other channel to perform forward and reverse TDT measurements. You measure the step transmitted through the device, rather than reflections from the device (as in TDR). Note: If the second channel is not connected to the same device as the first channel, crosstalk is displayed, as opposed to the step transmitted through the device. Take a 24. Take your measurement. For more information see Take measurement automatic measurements on page 25, or Take cursor measurements on page 26. 36 80E00 Electrical Sampling Modules User Manual