Download I-Max Touch
Transcript
SERVICE MANUAL Troubleshooting 8.3.3.8. DSPU IP address The DSPU has its own static IP address that is used for the connection with the external PC. The default set value for IP address is 010.000.000.215. 8.3.3.9. DSPU NET mask The DSPU has also a specific net mask, default set value of 255.255.255.0000. Using the standard method to change the value, set both according to values defined by the Network manager of the installation. * NOTE: Those values are used only by the DSPU, so only on the Digital units; do not update those in case of an analog (film) unit, and/or change those only to adhere at the specified values of the network manager in case of IP address conflict. I-Max Touch 8-38 (Rev. 2)