The optical signal of the detectors is converted by the microprocessor to a
numerical readout in conventional units of measure and displayed on the
front of the device. Other clinical parameters commonly used for the
assessment of oxygen and acid-base status are calculated from these
measured values.
6.5.4.3 The Reusable Sensor Cassette
The sensor cassette shown
without the syringe adapter
allows for direct sample asiration from a capillary or
Roche Microsampler. The
syringe adapter(s) are used
when sampling from standard
syringes.
Once removed from the
storage pouch the cassette
must be used within one
(1) our or be discarded.
Ca++
Outlet Port
Valve
Fillport
K+
Na+
Inlet Port
PCO2
PO2 / tHb/ SO2
pH
6.5.5 Calibration and Quality Control
6.5.5.1 Calibration
During the measurement and calibration processes, diagnostic tests are
automatically performed to assure correct operation of the instrument and
measurement of the cassette. These tests include automatic checks of the
cassette for integrity, temperature control, fluidic control during calibration,
proper equilibrium behavior of the sensors during calibration and
measurement, automatic detection of bubbles and short sample during
aspiration, and automatic detection of low gas, low battery, dirty optics, or
worn pump conditions.
Calibration of the tHb channel is required every 3 months. This calibration
is performed using the Roche tHb Calibration Cassette in a manner similar
to other instruments that measure tHb and/or hemoglobin derivatives
optically. The tHb calibration verifies the measurement optics and
electronics and corrects any potential drift. A second HbCal option is
available on the OPTI R. The HbCal – LastBlood option allows the OPTI R
to be correlated to another tHb measurement method. When the LastBlood
is run, the OPTI R will compensate for any measurement bias, allowing any
future measured samples to match the alternate device. For more
information, including detailed instructions, on the HbCal options, see
Section 6.3 “Quarterly Maintenance” in the OPTI R Operator’s Manual.