Download Pumps - SciCan
Transcript
96-104253 Rev. 2.0 Pumps Pumps Installing the Fluid-O-Tech Pump Assembly To install the pump assembly follow, these steps (see Figure 3): 1. Slide the pump assembly, inlet side of the pump away from the steam generator, into the mounting bracket and snap it into place. Do not kink the teflon™ tube (9). 2. Push the open end of the rubber inlet tube (6) extending from the water reservoir on to the pump inlet fitting (7) as far as it will go. Secure the tube to the fitting using a cable tie (4). 3. Attach the compression nut (8) on the Teflon™ tube to the inlet fitting (11) on top of the steam generator (10) finger tight. Tighten this nut using a 3/8-inch wrench. Do not overtighten. 4. Re-attach the diode plug (2) using the diode plug screw (1). 5. A dielectric strength test (Hi-Pot) and a protective bonding impedance test (ground continuity) must be performed on the Statim unit.See Required Information, Tools and Routine Maintenance. 6. Make sure there is sufficient steam-process distilled water in the reservoir and activate the pump using the Control Box for 5 seconds to ensure that the pump is functioning. 7. Fasten one cable tie to each end of the pump mounting bracket. 8. Run a sterilization cycle and observe all fittings and tubes for leaks. Check LCD read-outs for messages indicating cycle status. 9. Reinstall the cover. See STATIM Cover Removal and Replacement. 5 CONFIDENTIAL STATIM 1000 / 2000 / 2000S Cassette Autoclave Service Manual