10 • Maintenance and servicing
NE monitoring of critical
resistance
for
dual surfaced
neutral electrodes
Testing and measuring equipment
ERBE
Art. No.
20194-070
Description
20194-075
Patient cable NE
or
Patient cable NE, international
20100-033
Adapter cable NE
20100-101
VIO Testbox Symmetry/Resistance
(NE asymmetry/critical resistance)
20189-101
Dual-pedal footswitch with ReMode
Presets on the test specimen
• AUTO CUT, Effect 1, 10 watts.
• Neutral electrode "dual surface".
Test set-up and test procedure
Testbox 20100-101
ERBE
3
4
6
7
8
9
10
11
1
2
5
12
Fig. 10-18
• The test specimen is connected to the power supply via the power
cord.
• The test is performed without a load.
• The NE receptacle of the test specimen is connected to the VIO
Testbox via the patient cable NE with the adapter cable.
• The dual-pedal footswitch with ReMode is connected.
1. On the test specimen the (dual surfaced) neutral electrode lamp
must light up red.
2. Activate AUTO CUT via the footswitch. The test specimen must
inhibit activation and display or emit an optical and acoustic warning.
120 / 144
Art.-Nr.: 80116-271
04.06
1st test step