Download COL-1040
Transcript
4-6 4.6 “POWER LOW ERR” is Displayed Is the laser lit? Remove the rear cover on the main body and check that. YES A YES Calibrate the power meter. Is the output at the tip of the handpiece 40W or more (37W or more in the case of a scanner type)? (See 8.2.) NO Perform the power MAX test and adjustment of the CO 2 laser tube. (See 8.1.3) Is the problem solved? YES NO The troubleshooting is completed. YES See "6.2.3 Replacing the CO 2 Laser Tube". Make an adjustment to the master board. Is the problem solved? (See 8.2) YES The troubleshooting is completed. NO Does replacing the power meter sensor ASSY (EA41) solve the problem? (See 6.2.6.) YES NO Does replacing the power meter board ASSY (BA03) solve the problem? (See 6.2.5.) YES See "6.4.1 Replacing the Master Board ASSY (EA90)". YES The troubleshooting is completed.