Download R&S SMB100A Complete Manual
Transcript
Frequency Modulation - FM State - FM R&S SMB Activates/deactivates FM modulation. Activation of FM deactivates phase modulation. Remote-control command: SOUR:FM:STAT ON FM Source - FM Selects the source for the FM signal. The source can be selected independently for the different modulation types and the LF output. The configuration of the selected external and/or internal modulation source is performed in the lower section of the menu or in the Mod Gen block (internal source only). Int Selects the internal source for FM modulation. Remote-control command: SOUR:FM:SOUR INT Ext Selects the external source for analog frequency modulation. The external signal is input via the MOD EXT connector. Remote-control command: SOUR:FM:SOUR EXT Int + Ext Selects the internal and external source at the same time. This setting enables two-tone FM modulation. Remote-control command: SOUR:FM:SOUR INT,EXT FM Mode - FM Selects the mode for the frequency modulation. Normal The maximum range for modulation bandwidth and FM deviation is available.. Remote-control command: SOUR:FM:MODE NORM 1407.0806.32 4.66 E-2