Download R&S SMB100A Complete Manual
Transcript
R&S SMB Command Description SYSTem:SERRor? This command returns a list of all errors existing at the time when the query is started. This list corresponds to the display on the info page under manual control. The command is a query command and therefore has no *RST value. "SYST:SERR?" Example: 'queries all errors existing in the error queue. Response: "-221, 'Settings conflict', 153, 'Input voltage out of range'" 'the two returned errors have occurred since the error queue was last queried. *RST value Resolution - - Options SCPI Device-specific SYSTem:TIME 0...23,0...59,0...59 The command sets the time for the instrument-internal clock. SYST:TIME 12,0,0 Example: 'sets the time to precisely 12 pm. *RST value Resolution - - Options SCPI Compliant SYSTem:VERSion? The command queries the SCPI version with which the instrument complies. The command is a query command and therefore has no *RST value. SYST:VERS? Example: 'queries the SCPI version. Response: 1996 'the instrument complies with the version from 1996. *RST value Resolution - - 1407.0806.32 Options SCPI Compliant 6.97 E-2