Performance Verification 4. After the monitor completes POST, the monitor will: • be in SpO2 alarm • display an %SpO2 of 75 (pass criteria is 73 to 77 %SpO2 inclusive) • display a pulse rate of 60 (pass criteria is 57 to 63 BPM inclusive) • pulse amplitude indicator - display low level modulation (low amplitude pulse amplitude indicator) Test #1: BPM 1. Press the SRC-MAX % Pulse Rate selection button. The SRC-MAX Pulse Rate 200 LED will light. 2. The monitor BPM will increase to 200 and stabilize at 200 BPM. The test pass criteria is 197 to 203 BPM inclusive. 3. The monitor will display: 28 • 75 %SpO2 • 200 BPM (pass criteria is 197 to 203 BPM inclusive) • alarm • pulse amplitude indicator - low level modulation