Download PRISMA System
Transcript
Chapter 3: Continuous Renal Replacement Therapies (CRRT) Operation Startup Startup of the PRISMA Control Unit consists of the following steps: 1. Operator turns the power switch to the “on” position. 2. The control unit performs an initialization test to check the system electronics. The Logo screen is displayed, the non-mutable buzzer sounds, and all status lights are illuminated during the test. 3. When the initialization test is successfully completed, the Choose Patient screen appears on the display and the yellow status light illuminates. This indicates the PRISMA Control Unit is in the Setup mode and is ready for operation. Note: The above actions occur when a new PRISMA Control Unit is initially turned on. These actions also occur whenever the unit is turned on after being turned off in the Treatment Complete screen. If the control unit was last turned off in a screen other than Treatment Complete, a Query screen appears after the initialization test is completed. From the Query screen, the operator can choose one of two actions: • Begin on the same operating screen as when the unit was turned off (by pressing the CONTINUE key). • Start over at the Choose Patient screen (by pressing the RESTART key). Control and Navigation The PRISMA Control Unit is operated by means of the interactive display on the upper front panel. The screens displayed lead the operator through the operating procedures. Help screens provide additional information, if needed. The softkeys that appear on each screen enable the operator to give commands to the control unit and navigate between screens. ! WARNING PRISMA System Operator’s Manual If the display goes blank while power is on, immediately terminate the treatment and call for service. 45