Download Model 2000 Service Manual
Transcript
Table 6.4 Alarm and Alert Troubleshooting Alarm/ Alert Message Displayed Alarm/Alert Description System Status Troubleshooting 23 Patient Temperature 2 low. Check patient temperature. Patient Temperature 2 reading is below 32.0◦C (89.6°F) and water temperature is below 32.0◦C (89.6°F) and the system is unable to heat in Automatic Mode Changes to Stop Mode. Audio alarm half second tone every 10 seconds. Alarm/Alert indicator flashing Operate the system in Manual Mode with a setpoint above 32°C until the water temperature is greater than 32°C. Press Automatic and retest. . If the error persists, see General Troubleshooting - Will Not Heat. Patient Temperature 2 reading is equal or less than the userdefined value established in the custom settings menu. Current Mode continues. Audio alarm half second tone every 2 minutes. Alarm/Alert indicator flashing. Adjust Patient High Temperature Alert in the custom menu if desired. Patient Temperature 2 reading is above 38.0◦C (100.4°F) and water temperature is above 38.0◦C (100.4°F) and the system is unable to cool in Manual Mode Current Mode continues. Audio alarm half second tone every 2 minutes. Alarm/Alert indicator flashing. Operate the system in Manual Mode with a setpoint below 38°C until the water temperature is less than 38°C. If the error persists, see General Troubleshooting - Will Not Cool. Patient Temperature 2 reading is below 32.◦C (89.6°F) and water temperature is below 32.0◦C (89.6°F) and the system is unable to heat in Manual Mode. Current Mode continues. Audio alarm half second tone every 2 minutes. Alarm/Alert indicator flashing. Operate the system in Manual Mode with a setpoint above 32°C until the water temperature is greater than 32°C. If the error persists, see General Troubleshooting - Will Not Heat. Patient Temperature 2 reading is outside the lower limits of the display range in Automatic Mode. Changes to Stop Mode. Audio alarm half second tone every 10 seconds. Alarm/Alert indicator flashing. Use the CTU to verify the accuracy of the Control Module. Check connection of patient temperature probe and cable. If wet, replace cable. Check temperature cable connection at the rear of the Control Module. Flex the cable to ensure that there is not an intermittent fault. Test the input with a cable and probe proven good on another Control Module. If the Control Module proves inaccurate, replace the Isolation Card and retest. If the error persists, call Medivance Service. Calibrate upon completion. Alarm 24 Patient Temperature 2 below user defined limit. Alert 25 System is warming and Patient Temperature 2 high. Alert 26 System is cooling and Patient Temperature 2 low. Alert 27 Patient Temperature 2 probe disconnected or out of range. Alarm 47