TROUBLESHOOTING CHARTS (NOTE: Also see the RIB and Product Alert listing following these charts) PROBLEM Dark Slide Cycle with Front Door Open (Non-Continuous) (If cycling continuously go to Continuous Cycling chart) No Dark Slide Cycle PROBABLE CAUSE CORRECTIVE ACTION 1. Door pawl spring defective or not loaded under pack pawl. 1. Replace spring or load spring properly. 2. S1 slider fingers and switches on contact support block not interfaced properly. 2. Disassemble and install properly 3. S1 switch contacts deformed. (Closed with front door open - check with meter). 3. Re-form S1 contacts if possible. Replace contact support block if necessary. 4. Contact support block defective. 4. Replace contact support block. 1. S9 switch open (check with meter). support block if necessary. 1. Re-form S-9 switch contacts. Replace 2. S1 slider fingers and switches on contact support block not interfaced properly. 2. Disassemble and install properly. 3. S1 slider cam not mated to counter properly. 3. Disassemble and install properly. 4. S1 switch contacts deformed (closed with front door open - check with meter). 4. Re-form S1 contacts if possible. Replace contact support block if necessary. 5. Contact support block defective. 5. Replace contact support block. 6. Battery (+) wire not trapped in contact support block properly. 6. Install battery wire properly. 7. Flex circuit motor contacts not seated properly. 7. Install flex motor contacts properly. 8. Counter spring defective or not loaded properly. 8. Repair or replace spring. 79