CHECKOUT AND CONFIGURATION
Software Version 2.78
2.4
CONFIGURATION PROCEDURE
(Continued)
CAUTION
The instrument’s configuration information is
not lost when disconnected from AC power.
However, error history and infusion program
settings may be lost. If this information is to
be saved, be sure to record it before
disconnecting power.
2.4.1
Entering Configuration Mode
The instrument must be off (both channels
must be off for dual-channel).
The procedure for the single and dual
channel instrument is the same. Any
configuration in the dual channel sets the
same value for both channels.
1. Press and hold left-bottom display soft
key.
NOTE: For 7131/7231, the default code
will be 4EF3 (instead of BFFF) for
instruments set to factory defaults.
2.4.2
Setting to Defaults
The Set To Defaults mode programs all
configuration items to their default values.
CAUTION
It is strongly recommended that the
Configurable Options be reviewed for a
complete list of defaults before using this
feature.
1. Advance to C2 page.
A B
OPTIONS
Set To Defaults
Language
Air In Line
Dose Rate Drugs
page
2. Press POWER switch. Continue to press
display soft key until configuration mode
display appears, then release.
Page C1 of configuration mode is
displayed. This is a read-only display.
A B
OPTIONS
CONFIGURATION MODE
ID No. : 001234567
C
1
SW Rev.: 02.78
Config.: BFFF IVAC
C
2
2. Press Set to Defaults soft key.
A B
OPTIONS
Set All
configuration
C
2
items to their
default values
cancel
ok
page
3. Press page
configured.
to advance to page to be
3. Press ok to accept change and return to
beginning of C2 page. Pressing cancel
leaves all items set to their previous
values and returns to beginning of C2
page.
Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231
Technical Service Manual
2B-3