CORRECTIVE MAINTENANCE 5.3 TEST AND CALIBRATION (Continued) 5.3.2 Mechanism Visual Check (Continued) Figure 5-32. Mechanism Assembly Spring Location: Between motor plate and clamp arm (Partial sketched view of mechanism, for spring location reference only.) Spring connection to motor plate Back 4-Bar Figure 5-33. Mechanism Assembly Spring Location: Between back guide and clamp arm (Partial sketched view of mechanism, for spring location reference only.) Spring connection to back guide Back 4-Bar Spring connection to clamp arm (see exploded views below) Spring connection to clamp arm (see exploded views below) Correct installation Incorrect installation Exploded Views (spring connection to clamp arm) Correct installation Incorrect installation Exploded Views (spring connection to clamp arm) 5-40 Signature Edition® GOLD Infusion System, Models 7130/7131 and 7230/7231 Technical Service Manual