Download Model 3563 Integrating Nephelometer Operation and Service Manual

Transcript
The following chart represents some typical settings for the
previously mentioned time parameters.
Table 3-1
Typical Time Settings
Parameter
Ground Based
Ground Based (clean)
Airborne
Avg Time
Zero
Blanking
AZ Period
AZ Mode
60 to 300
300
30
3600
Normal
1500
300
30
720
Air Chop (SMZ = 5)
30
300
30
Manual
Manual
When using long averaging times (greater than 300 seconds), all
timing parameters should be set to integer multiples of the boxcar
size (see the STA command under "Set Commands" in Chapter 6).
This minimizes data loss that may occur if autozero transitions
occur during the middle of a measurement boxcar.
Operations Overview
3-9