Download CPB5800

Transcript
Pressure Balance
CPB5800
GB
3.2.3 Procedure for all dual-range piston-cylinder systems
.
3.2.3.1 Mass load
.
 Load the piston head with masses equivalent to the required pressure calibration point required.
Ensure the masses are correctly located in its respective spigot/recess.
Each mass has the following markings:
-Low Pressure/High Pressure Value
-Low Pressure/High Pressure Piston Area
-Mass set number
For high accuracy calibration, an additional marking (letter or letter/number
combination) is marked on the mass. This is to identify masses of similar nominal
pressure values, and thus obtain the actual mass value (grams) of said item.
 All dual-range piston-cylinder units have a basic head mass equivalent to 10 psi (on low pressure
area). If calibration is required in another pressure unit, the first mass applied to the piston head
should be the make-up mass (small mass with ‘+PISTON’ marking)
3.2.3.2 Approaching the pressure value
.
 The system must first be filled with oil and pre-compressed.
 For this the outlet valve must be closed.
 Operate the priming pump for several strokes. The pressure increases to a maximum of about 50
bar (depending on the volume of the connected test specimen).
 After that, increase the pressure by turning the built-in spindle pump clockwise.
 Just before the generated pressure reaches the actual calibration point, the masses should be
rotated by hand (approx 30-40 RPM) to ensure that the piston is in free-rotation. Care should be
applied when rotating the masses that no un-necessary transverse loads are applied to the piston.
Never rotate the piston-cylinder unit, if the piston is in the lower or upper block
position.
3.2.3.3 Pressure stable
.
 Continue generating pressure until the system is in a state of equilibrium.
 As the pressure calibration point is achieved, the piston will begin to move in an upward direction to
its ‘FLOATING’ position. On all dual range models, it has two ‘FLOATING’ (free rotation) positions
to correspond with the dual area piston unit.
One is for the low pressure area, and is indicated when a blue band with silver dashes becomes
visible.
One is for the high pressure area, and is indicated when a red band with silver dashes becomes
visible.
The bottom chamfered edge of the piston head floating anywhere within the above mentioned
bands indicates the piston unit is at pressure value of masses applied for the area it is operating
on. To confirm this, the operator can press down lightly (use index finger) onto the top of the
masses applied. If the piston and masses appear to bounce (move freely up and down) the piston
unit is at pressure value of masses applied.
WIKA Operating Instructions Pressure Balance Version 1.1
21