Load Cell Approvals
LC_Compatibility Worksheet instructions
AWG
Ø inches
Ø millimetres
area mm2
10
0.1019
2.59
5.26
11
0.0907
2.30
4.17
12
0.0808
2.05
3.31
13
0.0720
1.83
2.63
14
0.0641
1.63
2.08
15
0.0571
1.45
1.65
16
0.0508
1.29
1.31
17
0.0453
1.15
1.04
18
0.0403
1.02
0.82
19
0.0359
0.91
0.65
20
0.0320
0.81
0.52
14. Fraction of the maximum permissible error (mpe) for the connecting elements
(pcon). Unless otherwise specified in the NOA this value shall be 0.5. This field
has been disabled in this version of the worksheet.
15. Number of Divisions (n). This is a calculated field and must not be changed.
n = Max / e
B. Indicator (Ind)
1. Accuracy Class. Enter the Accuracy Class for the Indicating Element. This must
be a NAWDS approved Class and should be taken from the NOA when possible.
The Class entered in this field must reflect the device configuration. Entering any
other Class in this field is not acceptable.
2. Maximum Number of scale intervals (nind). Enter the maximum number of
intervals that the indicating element has been approved for. This information shall