Graphical User Interface (GUI)
may experience a boost in detector gain and laser power, an offset may be set
without the risk disappearing in the noise.
•
PCIS Range: The value entered here specifies the size of the window range
around the precursor. The default of 1% should not be changed. But, if this
value has been changed enter the precursor value a second time and press the
<Enter> key. flexControl calculates the new window range.
•
PCIS: The checkbox “On” is used to switch the precursor on or off. The button
opens the dialog Precursor Ion Selector Calibration (Figure 94).
Figure 94
Ion selector calibration dialog
If the Ion Selector has to be calibrated for a specific mass type it in the field
“Calibration mass”. The pulser timing will be calculated and has to be adjusted.
If a well fitting timing is found appy the button
. The new
and the constants
calibration can now be calculated with the button
are displayed in the field ‘Fit Results’. They are applied and transferred to the
field ‘Applied Calibratio Constants’ with the button
.
The button
•
top left on Figure 94 is used to reduce the size of this dialog.
LIFT: There is no direct access to a LIFT calibration if you work with a validated
LIFT method. The calibration can only be modified after invalidating the current
calibration with the button
. It is recommended to do a new calibration