Download User's Manual - X-Rite
Transcript
Positioning the Instrument on the White Calibration Standard . . . continued X-Rite Reflection Standard X-Rite Reflection Standard P/N SP68-62 Serial No. AXXXXXP-XXXXX Date XX/XX/XX WAVE [nm] REFL [%] 390 400 410 420 430 440 450 460 480 500 520 540 560 580 600 620 640 660 680 700 710 56.48 75.42 82.04 86.10 87.20 87.14 87.56 87.97 88.49 88.95 89.13 89.02 88.78 88.34 87.80 87.32 86.98 86.55 86.10 85.62 85.18 CALIBRATION CHECK D65/10 (WHITE) L* a* b* Instrument Model No. Instrument Serial No. COLOR CHECK Refer to section in operation manual appendix for procedure to obtain reference values. DATE: TEMP: ILLUM/OBS: APERT: REFERENCES VALUES STEP L* a* b* 1 (White) 2 (Blue) 3 (Brown) Notes: IMPORTANT! The standard is dramatically affected by smudge marks and dust; and must be kept clean. The Calibration Values for the white spot are affected by the environment and cleaning method of the standard. If the white spot does not measure correctly, it could be that the unit needs to be calibrated; there is dust in the optics; or the standard has smudge marks or is dirty. The ceramic standard should be cleaned using a mild soap and warm water solution, thoroughly rinsed with warm water, and wiped dry with a clean, lint free cloth. 3-3