Operation of the Instrument Installing and connecting HCLs You can use both single-element and multi-element hollow cathode lamps. Pb O PE M RA A T X E 12 1 0 Figure 6- 11 Installing a hollow cathode lamp 1. Open the lamp door. 2. Slide the required Lumina lamp into the lamp holder and make sure that the integrated socket slides fully home over the receptacle on the lamp holder. Align the lamp. 3. Close the lamp door. Installing and connecting EDLs If necessary, assemble the lamp by sliding the lamp driver into the lamp sleeve for the required element. Depress the locking pin as the sleeve slides into place and make sure that the locking pin extends through the locating hole in the sleeve. Single Coded EDL 1. Open the lamp door. 2. Insert the coding plug over the receptacle on the lamp holder. NOTE: Do not use coding plug for uncoded EDL. 6-22