Download AAnalyst 200 User's Guide
Transcript
AAnalyst 200 User’s Guide Disassembling the Burner Chamber These procedures assume that you have removed the burner assembly from the instrument and placed it on a workbench. Next you will disassemble it. 1. Press the safety latch tab on the ignitor box while gently pulling and twisting the burner head until it comes off the chamber. 2. Unscrew the retaining ring (Figure below) and pull out the burner o-ring. Figure 9- 7 Exploded view of the burner chamber assembly. 3. Lift the nebulizer clamp and carefully rotate the nebulizer arm clockwise (slightly) slide out the nebulizer. Put it to the side. 4. If you feel you’ll need to completely remove the end cap from the chamber, continue with steps 5 - 7. Otherwise, unscrew the four thumbscrews holding the burner end cap and carefully let the end cap hang down (you don’t have to disconnect any of the hoses). Now skip down to step 8. 9-25