Canister Change (in Carrying Case) cont.
®
8. Grasp the ActiV.A.C. Therapy Unit in one
hand and turn the unit and case so that the
canister is facing up. Continue to fold the
Access Flap back until the canister is fully
exposed.
and
8b
.
Canister
®
9a
9b
R
9. Position the ActiV.A.C. Therapy Unit so that
the Canister Latch Release can be pressed
toward the 50ml graduation mark.
See picture 9a .
D
8a
ev
See pictures
8b
8a
ib Ele
le ct
F r
N or oni
ot C c
Fo ur Pr
r C ren inte
om t R d D
m evi oc
er si u
ci on m
al
e
Pr Lev nt
in el
t
34
0
07
5
Press the Canister Latch Release and
remove the canister from the therapy unit.
See pictures 9a and 9b .
Dispose of the canister according to
institution and local environmental
regulations.
®
10a
10b
12a
12b
ns
10. Install the new canister onto the ActiV.A.C.
Therapy Unit by sliding the opening in the
canister over the Canister Latch Guide. Do
not twist or turn the canister as it is being
installed. See picture 10a .
rR
es
po
Push the canister onto the therapy unit. An
audible “click” should be heard.
See picture 10b .
U
se
18
11. Ensure the canister is securely installed on
the therapy unit. If the canister is properly
installed, the canister cannot be removed by
gently pulling the canister directly away from
the therapy unit.
Ensure that no part of the case is
between the canister and the
®
ActiV.A.C. Therapy Unit.
12. Pull the Access Flap back over the
®
canister and ActiV.A.C. Therapy Unit so
that the canister is covered by the case.
See pictures
12a
and
12b .
Access
Flap
Canister