Download PathwaysTM 4 Software User's Guide
Transcript
To adjust template points with a magnifier present, perform the following steps. 1 Deselect the Adjust Global option, and select Use Magnifier. 2 Move the pointer over each alignment point, and a magnifier window appears (an alignment hint picture also appears on the left panel). 3 Click anywhere in the magnifier window, and the alignment point snaps to this location. 4 The magnification can be increased or decreased using the up or down arrow keys while the magnifier is present. Perform the following steps to adjust alignment points without magnification:. Deselect the Use Magnifier check box. Click and drag the alignment point to move it. With ResGenTM GeneFilters microarrays, the template alignment points are located over 16 of the control spots for each microarray. In Field 1 of the GeneFilters microarray, the template alignment points are the upper right control spot in each grid. In Field 2 of the GeneFilters microarray, the template alignment points are the lower right control spot in each grid (refer to Appendix I). When the control spots are not visible (e. g., in cross-species hybridization), then the alignment points must be in the approximate position where the control spots normally reside; verify the accuracy of the alignment by viewing the overall template positioning. 70 RGMA10011 rev B