Download R&S FSVR Realtime Spectrum Analyzer User Manual

Transcript
R&S®FSVR
Remote Operation of the R&S FSVR
Remote Control – Commands
Suffix:
<n>
.
1...4
window
<ChannelType>
LEFT | RIGHt | MPX | MONO | STEReo | RDS | PILot
Channel type for which the command is performed.
Query parameters:
<MeasType>
ALL | ADEV | RDEV | SINad | THD | AFRequency | XTALk
ALL
All available measurement values
ADEV
The absolute deviation
RDEV
The relative deviation
SINad
The signal-to-noise-and-distortion value
THD
Total harmonic distortion
AFRequency
Audio frequency
XTALk
Crosstalk between left and right channels in dB:
-20|log(Left [kHz] / Right[kHz])| dB
*RST:
Return values:
<Result>
ALL
Measurement value according to the specified measurement type.
Note: if the specified result value is not available, a "Query
Error" is returned. If all result values are queried (query parameter
"ALL"), and one of them is not available, 9.91E+37 is inserted for
the missing value.
Example:
CALC1:MARK:FUNC:SFM:LEFT? THD
Queries the total harmonic distortion for the left stereo channel.
Usage:
Query only
Mode:
SFM
CALCulate:MASK Subsystem
The commands of the CALCulate:MASK subsystem configure the frequency mask trigger.
Programming example
TRIG:SOUR MASK
//Selects the frequency mask as a trigger source.
MMEM:MDIR 'C:\R_S\instr\freqmask\MyMasks'
CALC:MASK:CDIR 'MyMasks'
User Manual 1311.0687.32 ─ 04