Download Fluke 45
Transcript
Introduction and Specifications Specifications 1 Size 9.3 cm high, 21.6 cm wide, 28.6 cm deep Weight Net, 2.4 kg without battery; 3.2 kg with battery; Shipping, 4.0 kg without battery, 4.8 with battery. Power 90 to 264 V ac (no switching required), 50 and 60 Hz. 15 VA maximum Safety Compliant with the following standards: ANSI/ISA S82.01-1994 CAN/CSA-C22.2 No. 1010. 1-92 EN61010.1:1993 to 1000 V Overvoltage Cat I, 600 V Overvoltage Cat II UL3111-1. See the following for an explanation of CATI and CATII categories. CAT I: OVERVOLTAGE (Installation) CATAGORY I, Pollution Degree 2 per IEC1010-1 refers to the level of Impulse Withstand Voltage protection provided. Equipment of OVERVOLTAGE CATEGORY I is equipment for connection to circuits in which measures are taken to limit the transient over voltages to an appropriate low level. Examples include protect electronic circuits. CAT II: OVERVOLTAGE (Installation) CATAGORY II, Pollution Degree 2 per IEC1010-1 refers to the level of Impulse Withstand Voltage protection provided. Equipment of OVERVOLTAGE CATEGORY II is energy-consuming equipment to be supplied from the fixed installation. Examples include household, office, and laboratory appliances. Electromagnetic Compatibility Meets FCC Part 15 Subpart J. EN61326-1 (1998) RS-232-C Baud rates: 300, 600, 1200, 2400, 4800 and 9600 Odd, even or no parity One stop bit Options Battery (Option -01K) Type 8 V, Lead-Acid Operating Time 8 hours (typical). N lights when less than 1/2 hour of battery operation remains. Meter still meets specifications. 1-15