Download Fluke 45
Transcript
45 Service Manual by more than ±15%. After one second, the meter reverts to normal operation in calibration mode. Before suspecting a fault with the Fluke 45, verify that the calibration is being conducted properly. • Check the connections between the source and the meter. Are all the connections in place. • Check the output of the calibration source. Does it equal the value called for by this calibration step? • Check the calibration source. Is it in operate mode? Has it reverted to standby? If the calibration setup is correct, a faulty component within the Fluke 45 may be causing the failure. Each measurement function depends on a combination of components in and around the Analog Measurement Processor (A1U1). Basic dc measurements depend on the zener reference (A1VR1), reference divider network (A1Z3), and integrate resistors (A1Z3). Resistance measurements and dc measurements above three volts additionally depend on the resistors in the dc divider network (A1Z1). AC measurements depend on the ac divider network (A1Z2), ac buffer (A1AR1), and rms converter (A1A1), as well as the basic dc measurement components. Note The Fluke 45 utilizes three measurement rates: slow, medium, and fast. Two sets of measurement ranges are used: one for the medium and fast rates and one for the slow rate. Whenever the meter is powered-up or put in calibration mode, the medium rate is selected. During calibration, the slow rate is selected automatically as required by the calibration step. 5-18. Calibration-Related Components Measurement function calibration depends on various combinations of components in and around the Analog Measurement Processor. The major components are: • Rms Converter A1A1 • AC Buffer A1AR1 • Shunt Resistors A1R2, A1R3 • Zener Reference A1VR1 • Divider Network (DC/Ohms) A1Z1 • Integrate Resistors, Reference Divider A1Z3 • AC Divider Network A1Z2 • Rms Converter Network A1Z4 Calibration of the Fluke 45 utilizes a building block approach; individual components are calibrated separately and the appropriate calibration building blocks are assembled later for correction of any particular measurement function, range, and rate. Calibration steps are grouped by function, with dc volts being first. Therefore, the calibration constants are partitioned to allow for some independence in the correction of the various measurement functions. For example, if the zener reference and divider needed replacing, it would be necessary to recalibrate dc volts only; ac volts would not be affected. Table 5-6 identifies components that are unique to each calibration step. Table 5-7 provides a different approach, relating components that are common to a group of calibration steps. 5-20