Download User Manual
Transcript
8. MAINTENANCE Snifter Dust Monitor needs very little maintenance. To achieve maximum reliable operation the recommended maintenance interval is 2 months. Maintenance is done by removing the unit from the process and cleaning the sensor rod to prevent signal leakage to ground. If the particles in the gas are sticky and tend to build up, the cleaning needs to be done more often. Check the PC board is firmly connected inside the enclosure. Check LED indicator cable and signal/power cable connectors are connected correctly. 9. TROUBLESHOOTING 9.1. 1. 2. 3. The Monitor is not giving relay output signal Check the power and signal wiring are connected correctly. Check there is power on (cover LED glows). Do the auto setup procedure (green LED starts blinking – stops when ready). If the monitor is not giving any relay output signal after checks 1, 2 and 3 contact your local distributor. 9.2. 1. 2. 3. The Monitor is not responsing after autosetup procedure Check that there is normal process going on and there were normal operation conditions during auto setup. Check the power and signal wiring are connected correctly. Check the signal is not leaking to ground. - There is NO contact between metal sensor probe and duct wall. - Ensure the gas is not condensing (causes signal leakage) - Check the sticky dust does not build up on the base of the sensor and no bridging between sensor probe and duct wall. 13