Performance Monitoring Use Your Test Method (Known Performance) Plate Count - Efficiency of the Separation * Monitor at least One Peak in one injection - Plate Count (Peak width relative to RT), - Peak Asymmetry, - Retention Time and/or Retention parameter - Relative Retention Time for Critical Pair of Analytes. - Peak Response * A "Plate Count" Actually Is a Determination Of Both The Column AND Instruments' Performance * Inject Multiple Runs - Precision (at least 5 injections) - Accuracy (Use Control Samples) Performance Monitoring Column Efficiency: N = the number of Theoretical Plates a = is a constant depending on the Method used tr = retention time of peak W = the peak width (time units) at a given peak height Performance Monitoring Band Spreading tR N=a ( tr W 2 ) h t0 w METHOD Peak Width at Half Height Peak Width at 4.4% Peak Height (5 Sigma) Tangent a 5.54 25.0 16.0 Dr. Shulamit Levin, Medtechnica * Band Spreading Impacts Chromatographic Performance -- The Greater The Band Spreading, The Poorer The Performance (ie; Resolution) * Band Spreading Contains Both An Instrument AND A Column Contribution 2