Download LabChip GX/GXII User Manual
Transcript
Software Reference 210 Optics Calibration Window Use the Optics Calibration window to calibrate the instrument using the Test Chip C as the calibration standard. (If the CFR Support option is installed, only displays if the user has Perform Validation access right.) To open the Optics Calibration window, select Tools Calibrate Optics on the LabChip GX Main Window. Figure 89. Optics Calibration Window The Optics Calibration Window contains the following options and buttons: Reference Contrast - Contrast value from Test Chip C that will be compared to the scan results Contrast. Instrument Gain Correction - The current correction factor for the instrument. Chip Temperature (Target 35) degC - The current temperature of the chip. The chip is always warmed to the target temperature to more accurately calibrate the optics. Background - The background measured on the current instrument after focusing. Contrast - The contrast measured on the current instrument. New Correction Factor - The correction factor calculated from the instrument (Reference Contrast / Scan Results Contrast). V3.1 LabChip GX User Manual Caliper Life Sciences, Inc.