21 5.2.3 Trigger auto tab This tab allows the modification of the default parameters for advanced recognition. Refer to § Chapter 7: Automatic measurement type recognition for more information. Please note that the modification of these parameters can alter the automatic recognition process; it is therefore necessary to perfectly understand the mechanism before making any change. Event # : 5 events are activated and cannot be deleted Code : 5 codes are defined; one for each event Fast logging: always Yes Logic for triggers: And or Or; if several triggers are defined Trigger # : number of the trigger Trigger type : selection of the data to be used for trigger Start trigger [dB]: level for trigger start End trigger [dB]: level for trigger stop Pre-trigger [Sec]: buffered time before trigger starts Post-trigger [Sec]: minimum delay before trigger stops Minimum time duration [Sec]: minimum duration for a trigger to start End duration [Sec]: time duration after end trigger level to stop the trigger 5.2.4 Signals tab Audio signal sampling frequency [kHz]: selection of the audio sampling frequency; selectable among 51.2; 25.6; 12.8; 6.4; 3.2 and 1.6 kHz. Default value is 12.8 kHz Audio recording with measurement: No or Yes; default is Yes (only if option A is active) Brand of ACOEM