Download Operators Manual
Transcript
Front Panel Operation Sample Applications 4 4. Test the ac voltage function: a. Press R on the calibrator and set the DMM function switch to K. b. Set the output of the calibrator to 350 mV at 60 Hz. and press O. Verify the errors are within specifications. c. Check the error against specifications at the following voltages and frequencies: Voltage Frequency 350 mV 60 Hz, 5 kHz, & 20 kHz 3.500 V 60 Hz, 5 kHz, & 20 kHz 35.00 V 60 Hz, 5 kHz, & 20 kHz 329.0 V 60 Hz, 5 kHz, & 20 kHz 100.0 V 20 kHz 200.0 V 20 kHz 300.0 V 20 kHz 1000 V 60 Hz & 5 kHz 5. Test the Frequency function: a. Press Ron the calibrator, set the DMM function switch to p, and press Hz on the DMM. b. Set the calibrator to 150 mV at 19.0 kHz and press O. Verify the error is within specification. c. Set the calibrator to 150 mV at 190 kHz. Hint: press e twice to move the cursor to the frequency reading in the output display and press X. Verify the error is within specification. 6. Test Frequency Sensitivity and Trigger Levels: a. Press R on the calibrator, set the DMM function switch to pand press Hz on the DMM to choose the frequency mode. b. Set the calibrator to 300 mV at 1 kHz and press O. Verify the frequency error is within specification. c. Change the calibrator output to 1.7 V. Verify the frequency error is within specification. d. Change the calibrator output to 1.0 V. Verify that the DMM displays 000.0 frequency. e. Change the DMM range to 40 V by pressing RANGE. Change the calibrator output to 6.0 V. Verify the frequency error is within specification. f. Change the calibrator output to 2.0 V. Verify that the DMM displays 000.0 frequency. 4-63