Download Operators Manual

Transcript
5520A
Operators Manual
FUNC?
x
IEEE-488
x
RS-232
x
Sequential
x
Overlapped
x
Coupled
(Function query) Returns the present output, measurement, or calibration function. See
the response below for output & measurement modes.
Responses:
DCV
ACV
DCI
ACI
RES
CAP
RTD
TC_OUT
DC_POWER
AC_POWER
DCV_DCV
ACV_ACV
TC_MEAS
SACV
SDCV
MARKER
LEVSINE
EDGE
(dc volts function)
(ac volts function)
(dc current function)
(ac current function)
(ohms function)
(capacitance function)
(temperature with an rtd function)
(temperature with a thermocouple function)
(dc power function)
(ac power function)
(dual dc volts function)
(dual ac volts function)
(measure temperature with a thermocouple)
(oscilloscope ac volts function)
(oscilloscope dc volts function)
(oscilloscope marker function)
(oscilloscope leveled sine function)
(oscilloscope edge function)
Example:
FUNC? returns DCV_DCV
Returns DCV_DCV when the Calibrator output function dual dc volts.
HARMONIC
x
IEEE-488
x
RS-232
x
Sequential
x
Overlapped
x
Coupled
(Harmonic command) Makes the frequency of one output a multiple of another output
for the ac voltage or ac power functions (sine waves only). For example, in dual ac
voltage, have the frequency of the voltage output on the Calibrator front panel
NORMAL terminals at 60 Hz and the frequency of the voltage output on the AUX
terminals at the 7th harmonic (420 Hz). The range for the harmonics is 1 to 50.
Parameters: <value>, PRI
<value>, SEC
Example:
(fundamental at 5520A NORMAL terminals)
(fundamental at 5520A AUX terminals)
HARMONIC 5, PRI
Load the fundamental frequency at the primary (PRI) output (NORMAL terminals), and
the 5th harmonic frequency is at the secondary output (AUX terminals). For example, if
the fundamental frequency output is 60 Hz, the harmonic frequency output is 300 Hz.
6-20