Download User manual AS110, edition 7
Transcript
Chapter 2 39 The AS 110 standard is fitted with the following tubing: Tubing Materials/Dimensions Standard sample needle and tubing (label 15 µL) SS: 97 mm x 0.8 mm OD x 0.25 mm ID ETFE (Tefzel): 200 mm x 1/16” OD x 0.25 mm ID Buffer tubing from high-pressure valve to syringe valve (label 1000 µL) ETFE (Tefzel): 1275 mm x 1/16” OD x 1.0 mm ID Syringe valve to wash solvent bottle PTFE: 400 mm x 1/8” OD x 1.6 mm ID Syringe valve to waste PTFE: 400 mm x 1/8” OD x 1.6 mm ID The AS 110 micro is fitted with the following tubing: Tubing Materials/Dimensions Standard sample needle and tubing (label 2.4 µL) SS: 136 mm x 0.8 mm OD x 0.1 mm ID PEEKsil: 165 mm x 1/32” OD x 0.1 mm ID Buffer tubing from high-pressure valve to syringe valve (label 50 µL) PEEK: 260 mm x 1/16” OD x 0.5 mm ID Syringe valve to wash solvent bottle PTFE: 400 mm x 1/16” OD x 0.75 mm ID Syringe valve to waste PTFE: 400 mm x 1/8” OD x 1.6 mm ID The AS 110 micro is fitted with the following tubing: Tubing Materials/Dimensions Standard sample needle and tubing (label 2.4 µL) SS: 136 mm x 0.8 mm OD x 0.1 mm ID PEEKsil: 165 mm x 1/32” OD x 0.1 mm ID Buffer tubing from high-pressure valve to syringe valve (label 50 µL) PEEK: 260 mm x 1/32” OD x 0.5 mm ID Syringe valve to wash solvent bottle PTFE: 400 mm x 1/16” OD x 0.75 mm ID Syringe valve to waste PTFE: 400 mm x 1/8” OD x 1.6 mm ID