Download TraceFinder 3.2 User Guide Optimized for

Transcript
5
Using the Method Development Mode for Quantitation Methods
Editing a Master Method
Figure 42. XIC Filter dialog box
Table 24. XIC Filter dialog box parameters (Sheet 1 of 2)
Parameter
Description
Mass Analyzer Any: Allows any mass analyzer.
Type
FTMS: Fourier Transform Mass Spectrometer
ITMS: Ion Trap Mass Spectrometer
Sector: Static electric or magnetic sectors, or a combination of the two
SQMS: Single Quad Mass Spectrometer
TOFMS: Time-of-Flight Mass Spectrometer
TQMS: Triple Quad Mass Spectrometer
178
TraceFinder User Guide
MSX
Any: Allows both MSX and non-MSX scans.
On: Allows only MSX scans.
Off: Allows only non-MSX scans.
Data
Dependent
Any: Allows both data-dependent and non-data-dependent filters.
On: Allows only data-dependent filters.
Off: Allows only non-data-dependent filters.
MS Order
Any: Allows any MS order.
MS: Single mass spec stage
MS2-MS3: Multiple mass spec stages
Polarity
Any: Allows both positive and negative.
Positive
Negative
Thermo Scientific