7
Using the Acquisition Mode
Sample Types
Sample Types
The TraceFinder application uses the following sample types in all sample definitions and
reports.
Table 76. Sample type definitions
Thermo Scientific
Sample type
Definition
Negative
Contains no target compounds but might contain an ISTD when you
use the internal standard quantitative analysis technique. By analyzing a
blank sample, you can confirm that there are no residual compounds in
the solvent system that can cause erroneous results.
Unextracted
Similar to a Negative sample, but contains target compounds. By
analyzing a sample, you can confirm that there are no residual
compounds in the solvent system that can cause erroneous results.
Calibrator
(Calibration standard) Contains known amounts of all target
compounds. The purpose of a standard is to measure the response of the
instrument to the target compounds so that the processing software can
generate a calibration curve for each compound.
QC
(Quality Check) Contains a known amount of one or more specific
target compounds. The application places check standard samples in the
sequence so that it can test quantitative analysis results for quality
assurance purposes. After the application analyzes the QC sample, it
compares the measured quantity with the expected value and an
acceptability range. The quantitative analysis of a QC sample is
classified as passed if the difference between the observed and expected
quantities is within the user-defined tolerance. A QC sample is classified
as failed if the difference between the observed and expected quantities is
outside the user-defined tolerance.
Solvent
Contains only solvent.
Specimen
Used for quantitative analysis of samples.
Hydrolysis
Checks the degradation of compounds dissolved in water.
TraceFinder User Guide
367