BA-03X User Manual _________________________________________________________________________________________________________________ CURRENT RANGE unit The CURRENT OUTPUT unit consists of (8) CURRENT RANGE LED and (9) CURRENT RANGE switch (8) CURRENT RANGE LED The CURRENT RANGE LED indicates the setting x10 (±120 nA) of switch (9) (LED: ON). (9) CURRENT RANGE switch Switch for setting the CURRENT RANGE of the amplifier: x1: max. ±12 nA into 1 G, x10: max. ±120 nA into 100 M HOLDING CURRENT unit The HOLDING CURRENT unit consists of (10) HOLDING CURRENT (nA) potentiometer and (11) + / 0 / - switch (holding current switch). (10) HOLDING CURRENT (nA) potentiometer With this control a constant current (holding current) can be generated (ten turn potentiometer, clockwise), calibrated in nA (Imax = 10 nA or 100 nA, dependent on setting of CURRENT RANGE switch #9). The polarity of the holding current is set by toggle switch (11). (11) + / 0 / - switch (holding current switch) Switch to disable holding current generation or to set the polarity of the holding current (+: current positive, 0: holding current disabled, -: current negative). (12) BIAS (bias current) potentiometer 10 turn potentiometer for setting the output current of the headstage (headstage BIAS current) to zero (ten-turn potentiometer, symmetrical, i.e. 0 pA = 5 on the dial), range: 200 pA (CURRENT RANGE: x1), (see chapter 7.1). ___________________________________________________________________________ version 2.0 page 10