Download OQ and PQ Operating Instructions
Transcript
OQ and PQ Operating Instructions Position of the sensor tip behind the capillary clip Figure 4: ACC-3000(T) – Position of the temperature sensor 3.3.2 Configuration RF2000 fluorescence detector Before you connect the RF2000 fluorescence detector with Chromeleon, set the ZWAVE parameter to 1. You can set this value only on the instrument: Disable the keyboard interlock by simultaneously pressing <Shift> and <CE>. Press <func> until the ZWAVE command appears on the display. Press <1> on the number keypad and confirm with <Enter>. Enable remote operation. To do so, press <func>, until the RS232 command appears on the display. Confirm with <Enter>, and then press <func>. On the display, the reading is CONNECT. Confirm with <Enter>. You can then connect the instrument with Chromeleon. AD25 UV detector On the Signals page for the AD25 in the Server Configuration program, change the unit to mAU (instead of AU) and the factor to 1000 (instead of 1.000). WPS-3000(T)PL autosampler For successful qualification of the Dionex WPS-3000(T)PL autosampler, make sure that the Upgrade Kit for a 250 µl-syringe is installed. WPS-3000TBPL Analytical autosampler In order to ensure a successful qualifcation, the Dionex WPS-3000TBPL Analytical autosampler must be equipped with the standard or large volume configuration. In addition, it must be activated in the Server Configuration on the Options page of the "WPS-3000TBPL Analytical“ driver: HPLC_OQ_PQ_E_Manual.doc – Version: 6.8 dated Oktober 2008 Page 31 of 95