Download OQ and PQ Operating Instructions

Transcript
OQ and PQ Operating Instructions
Position of the sensor tip behind the capillary clip
Figure 4: ACC-3000(T) – Position of the temperature sensor
3.3.2 Configuration
 RF2000
fluorescence
detector
Before you connect the RF2000 fluorescence detector with Chromeleon, set the ZWAVE
parameter to 1. You can set this value only on the instrument:
 Disable the keyboard interlock by simultaneously pressing <Shift> and <CE>.
 Press <func> until the ZWAVE command appears on the display.
 Press <1> on the number keypad and confirm with <Enter>.
 Enable remote operation. To do so, press <func>, until the RS232 command appears on the
display. Confirm with <Enter>, and then press <func>. On the display, the reading is
CONNECT. Confirm with <Enter>. You can then connect the instrument with Chromeleon.
 AD25 UV detector
On the Signals page for the AD25 in the Server Configuration program, change the unit to mAU
(instead of AU) and the factor to 1000 (instead of 1.000).
 WPS-3000(T)PL autosampler
For successful qualification of the Dionex WPS-3000(T)PL autosampler, make sure that the
Upgrade Kit for a 250 µl-syringe is installed.
 WPS-3000TBPL Analytical autosampler
In order to ensure a successful qualifcation, the Dionex WPS-3000TBPL Analytical autosampler
must be equipped with the standard or large volume configuration. In addition, it must be activated
in the Server Configuration on the Options page of the "WPS-3000TBPL Analytical“ driver:
HPLC_OQ_PQ_E_Manual.doc – Version: 6.8 dated Oktober 2008
Page 31 of 95