Download MICROX TX3 - Loligo Systems
Transcript
Calibration 59 6. Enter the ‘actual atmospheric pressure’. The atmospheric pressure of the calibration is needed to convert the oxygen unit % air-saturation into partial pressure units (hPa, Torr) or concentration units (mg/L, µmol/L). Please note that changes in the actual atmospheric pressure have no effect on the partial pressure units (hPa, Torr) and concentration units (mg/L, µmol/L), but the oxygen units % air-saturation and % oxygensaturation have to be corrected for air pressure changes. 7. Place the vessel with the label ‘cal 0‘ underneath the microsensor. Please ensure that the sensor tip is in the protective glass housing. Locate the glass housing carefully about 5 mm above the water surface. Ensure that the temperature sensor has been dipped about 1-2 cm into the calibration solution. 8. The glass fiber with its sensing tip is prevented from slipping using a protection tubing. Slacken the protection tubing from the glass housing, extend the sensor tip about 1 cm from the glass housing and fix the glass fiber again with the protection tubing. . WHEN THE GLASS-FIBER WITH ITS SENSOR TIP IS PUSHED OUT, HANDLE WITH CARE. THE GLASS FIBER IS UNPROTECTED AND MIGHT BREAK! 9. Ensure that the sensor tip is dipped about 4 mm into the calibration solution 0, but not the protective glass housing.