Transcript
72 Trouble Shooting 11 Trouble Shooting Error Cause Action Switch on device with ON/OFF switch on the rear panel Connect power supply with device Device does not work and LED on Device is not switched on the front panel is not lit No power supply Device does not work and LED on No connection to PC the front panel is on Check connection of the device to your PC (RS 232) Temperature compensation failed, PT 1000 sensor is not connected Check connection, no temperature measurement properly possible PT 1000 sensor is faulty Contact our service Warning light: Amplitude: red Microsensor properly is not connected Check connection connector of the ST- ST-connector is contaminated Clean connector with a soft, lint-free cloth rag Sensor tip is damaged Replace oxygen sensor; (send the sensor back to PreSens for recoating service) Warning light: Phase: red Phase angle out of limits Check connection of the microsensor Replace microsensor Warning light: overload: red Too much false light Reduce false light Use optically isolated microsensors No cal Calibration failed Calibrate again Check calibration solutions Sensor was not in the right calibration Immerse Sensor standard calibration vessel Strong signal fluctuations in the proper Sulfite solution has aged Prepare fresh sulfite -solution Air bubbles at sensor tip The glass fiber with its sensor tip is not extended from the syringe needle Amplitude is too low Remove air bubbles by carefully tapping Completely extend the sensor Check amplitude