Download PerkinElmer Perkin Elmer - lambda 20 - Perkin Elmer Uvvis-spectrophotometer Lam
Transcript
Error Messages Error No points stored. Start measurement. Meaning The method was interrupted with [STOP] before data could be stored, for example, during the equilibrium time of a substrate method. Restart the method. Two solutions to 1 ABS value. Change curve fit algorithm. In a non-linear calibration curve two reference solutions exhibit the same absorbance value. Check the references, or change the concentration range, or change the curve fit algorithm. Value not within valid limits. Check references or change curve fit alogrithm. The measured sample concentration is outside the calibration range. Measure additional calibration solutions within the concentration range or change the curve fit algorithm. Wavelength data do not fit. Start background correction. Appears with TIME DRIVE methods if the wavelength has been modified since the last background correction. Perform a new background correction using the proper method. 165